Error code E-1012
Probe not detected
Error code E-1012 means a connected ultrasound probe is not detected or is reported as incompatible. Often the cause is not the probe itself but the connector, the cable or the probe interface board in the system. We pinpoint the cause precisely and repair at component level.
What the message means
The system fails to detect a connected ultrasound probe or reports it as incompatible.
Typical symptoms
This is how you recognise the fault in operation.
- Connected probe is not detected or only intermittently
- "Probe incompatible" message despite a matching probe
- Detection drops out when the cable or connector is moved
- A single probe port fails while others still work
Possible causes
This is what experience shows is behind it.
- Faulty probe connector or bent contact pins
- Broken conductor in the probe cable or at the strain relief
- Fault on the probe interface board in the system
- Corrosion or contamination on the connector contacts
Before / after
The fault on the left, the same system after our repair on the right.
A common cause when the probe is not detected: cable and connector damage. On the left the damage, on the right the restored probe from our workshop.
How we proceed
Diagnosis
We cross-test the probe on a known-good port and a known-good probe on the affected port. This unambiguously separates whether the probe, the connector or the probe interface board in the system is the cause. That is the basis for a precise repair.
Repair
We repair faulty connectors and cables on both probe and system; contact pins are straightened or replaced. If the fault is on the probe interface board, we restore it at component level instead of swapping it. Including a function test of all probe ports, a test report and a warranty, usually within 1 to 2 working days.
Affected systems
This error code is documented on 12 systems. Go to the device-specific fault page:
Frequently asked questions about this error code
Not necessarily the probe. Very often the fault sits in the connector, the cable or the system's probe interface board. We cross-test probe and port and isolate the cause clearly before anything is replaced.
Yes. After the repair we test all probe ports for reliable detection, so a second port doesn't develop the same fault.
We fix E-1012.
Send us a photo of the screen, ideally via WhatsApp. We'll identify the fault, give you a realistic cost estimate upfront and tell you what to do.
A photo says more than a thousand words.
Send us a photo of the screen with the message, ideally via WhatsApp. That way we match the fault to the component right away and get back to you with a diagnosis and a fixed price, straightforward and personal.
We provide a suitable loan unit from our stock, matched to your system. Just get in touch and we'll arrange everything personally.
All error messagesIncluded with every repair